The i2000 is an AC current probe utilizing the Rogowski principle. It can be used to measure currents up to 2000A when used in
conjunction with multimeters, recorders or data loggers. The flexible and lightweight measuring head allows quick and easy installation in
hard to reach areas.
Three ranges: 20A, 200 A & 2000 A, 100 mV/A, 10 mV/A & 1 mV/A output
Ideal for current measurement on thick and hard to reach conductors without breaking the circuit
Maximum conductor diameter 178 mm
CAT III 600V safety rating
Electrical specifications
-Current ranges 20 A; 200 A; 2000 A AC rms
-Output sensitivity (AC-coupled) 100 mV; 10 mV ; 1 mV; 1 mV per A
-Load impedance 100 kOhm minimum
-Accuracy at +25 °C ± 1 % of range (45 Hz to 65 Hz)
-Linearity 10 % to 100 % of range ± 0.2 % of reading
-Noise 8 mV ac rms (30 A) 2 mV ac rms (300 A, 3000 A)
-Frequency range 10 Hz to 20 kHz (-3 dB)
-Phase error < ± 1 º (45 Hz to 65 Hz) ± 10 º (at 20 kHz)
-Position sensitivity with cable > 25 mm from the coupling ± 2 % of range
-External field with cable > 200 mm from the head ± 1 % of range
-Power supply 2 x AA/MN 1500/LR6 Alkaline, 200 hours, low battery indicator
-Temperature Coefficient ± 0.08% of reading per°C
-Working voltage see Safety standards section 600 V AC rms or DC
General Specifications
-Head Circumference 610 mm (24 in)
-Head cable diameter 14.3 mm (0.56 in)
-Head bend radius 38.1 mm (1.5 in)
-Cable length (head to electronics) 2 m (6.56 ft)
-Output connection Molded 4 mm safety plugs (not integrated)
-Operating temperature range 0 °C to +70 °C (32 °F to 158 °F) (head and electronics)
-Storage temperature range -40 °C to +105 °C (-40 °F to 221 °F) (head) -20 °C to +85 °C (-4 °F to 185 °F) (electronics)
-Operating humidity 15 % to 85 % (non-condensing)
-Weight 180 g (6.34 oz) (head) 190 g (6.70 oz) (electronics)
-Safety standards EN 61010-1: 2001; EN 61010-2-032: 2002; EN 61010-031: 2002; 600 Vrms, Category III, Pollution Degree 2 U-Use
Use of the probe on uninsulated conductors is limited to 600 V ac rms or dc and frequencies below 1 kHz.
-EMC standards EN 61326: 1998 +A1, A2, & A3
-Dimensions HxWxD 116 mm x 68.5 mm x 30 mm (4.56 in x 2.69 in x 1.18 in)